In two con gurations, a solid-state acousto-optic (AO) de ector or modulator is mounted in a 0.5 m monochromator for background correction with inductively coupled plasma atomic emission spectrometry (ICP-AES). A fused silica acousto-optic modulator (AOM) is used in the ultraviolet (UV) spectral region applications while a glass AO de ector (AOD) is used for the visible (VIS) region. The system provides rapid sequential observation of adjacent on- and off-line wavelengths for background correction. Seventeen elements are examined using pneumatic nebulization (PN) and electrothermal vaporization (ETV) sample introduction. Calibration plots were obtained with each sample introduction technique. Potable water and vitamin tablets were analyzed. Flame atomic absorption (FAA) was used to verify the accuracy of the AO background correction system.
Publication Date
1-1-2003
Recommended Citation
Miller, H. M.; Spudich, T. M.; Carnahan, J. W. Development and application of acousto-optic background correction for inductively coupled plasma atomic emission spectroscopy. (2003) Appl. Spectrosc., 57: 703-710.
Original Citation
Miller, H. M.; Spudich, T. M.; Carnahan, J. W. Development and application of acousto-optic background correction for inductively coupled plasma atomic emission spectroscopy. (2003) Appl. Spectrosc., 57: 703-710.